one·Research on Vacuum Growth Furnace for Crystal Growth Process Equipment purpose
equipmentMainly used forScientific research on crystal growth process can be used to prepare oxide single crystals and metal single crystalscrystalPreparation, such as sapphireGGG, YAG, LaAlO3, Si, Ge, etc
two·Research on Vacuum Growth Furnace for Crystal Growth Process Product Features
1·Modular design, compact structure, beautiful appearance,
2. AdoptingSide door openingExquisite structure, convenient, smooth movement, easy loading and unloading of materials
3. Adopting a touch screen+PLC control method, it has a high degree of automation, intuitive operation, and powerful functions.
4. Dozens of sintering processes can be pre stored on the touch screen, which can be edited once and directly called for later use, saving the trouble of multiple editing processes, avoiding input errors, and burning out products.
5. The sintering temperature, vacuum degree, and other data can be recorded in real-time, and can also be started and stopped at any time to reduce useless data. The data can be queried and exported for download.
6. High sintering temperature, using graphitecrucible, reachable2100℃,
three·Equipment parameters
1. Total power of equipment:≤35Kw; Power supply voltage: 3-phase 380V, 50Hz
2. Design temperature:2100℃ heating power ≤25Kw three-phase380V 30-80KHz high-frequency induction heating
3. Rated temperature: room temperature~2100℃
4. Induction coil size:F90×100Mm (diameter x height),) Set an insulation layer in the middle and place the bottom of the crucible inside the coil to prevent the bottom material from solidifying.
6·Chamber size horizontalF400F500mm(Temporary)
7Temperature measurement system:Design temperature measurement holes on the side of the crucible Configure infrared thermometer and Eurofins programming instrument
9. liftspeed Slow lifting and lowering0.1-10Mm/h (servo control)
10. Pulling rod rotation speed0.1-25r/min
11. Rapid lifting speed of lifting rod:≤35mm/min
12. Lifting itinerary:≤300mm
13. Furnace inflation pressure:≤0.03MPa
14. Cold state ultimate vacuum degree:6.7*10-4PA (equipped with vacuum gauge, vacuum gauge tube equipped with oil ceramic anti-corrosion coating)