The Japanese NOISEKEN 2KV high-frequency noise simulator NS-S220 meets the Japanese JIS industrial standard, with a high voltage of 2kV and a bandwidth of up to 1 μ s. The new model uses semiconductor relays to improve the stability of the test pulse waveform and adds options for various tests.
NOISEKEN 2KV High Frequency Noise Simulator from JapanINS-S220
Meets Japanese JIS industrial standards, with a high voltage of 2kV and a bandwidth of up to 1 μ s. The new model uses semiconductor relays to improve the stability of the test pulse waveform, enabling better quantitative and reproducible testing than before; Add options for various tests.
NOISEKEN 2KV High Frequency Noise Simulator from JapanINS-S220Product Features:
Can achieve high-speed pulse cycle switching testing: because the frequency of pulse switching is very fast, the occurrence rate of misoperation will also increase, and it can be expected that the testing time will be shortened;
The setting of pulse width is very simple: simply press the button to set in 50ns steps, which shortens the setting time and reduces the occurrence of wiring errors;
The stability of the test pulse waveform is better, and it can achieve highly reproducible testing;
Reduced the quantity of consumables and effectively lowered usage costs;
By using a short-circuit plug, it is very easy to switch between common mode/differential mode testing;
The instrument is equipped with a 50 Ω terminal, which is easy to wire;
Through the output panel, the EUT's AC plug can be directly connected; (Optional)
Simulators can conduct various experiments through various probes, coupling clips, adapters, etc; (Optional)
By adding a coupling network, experiments on EUT power supply three-phase 5-wire can be conducted; (Optional)
Specifications:
project |
Specifications |
Pulse setting range ① |
Pulse output voltage |
0.5kV~0.99kV ± 10% 0.01k step |
pulse width |
100ms~1000ns±10% |
repetition cycle |
1ms~999ms ± 10% 1ms step |
Pulse setting range ② |
Pulse output voltage |
1.00kV~2.00kV ± 10% 0.01kV step |
pulse width |
50ns~1000ns ± 10% 50ns step |
repetition cycle |
10ms~999ms ± 10% 1ms step |
output voltage |
0.5-2kV ± 10% (10V step) |
polarity |
Positive/Negative |
rise time |
Below 3ns |
output impedance |
50Ω |
Terminal impedance |
50Ω |
repetition cycle |
Line phase |
50Hz/60Hz injection phase angle 0-360 °± 10 ° (1 ° step) superimposed with L-N synchronous output
|
duration |
1ms to 999ms ± 10% (pulse setting range ①)
10ms to 999ms ± 10% (pulse setting range ②)
|
external trigger |
Selection period: 10ms or more Pulse width: 1ms or more Input level: TTL/Open circuit rectifier negative logic
If LINE PHARE mode is selected and the input period is 16-20ms, it will be recognized as the synchronization signal of CDN
|
1 SHOT |
Press the 1SHOT key once to output once
Synchronize the phase angle output with the phase angle output in PHARE mode
|
Memory capacity |
5 experiments |
Test time |
1-999s ± 10% 1-second step |
Overlay phase switching |
L (+), N (-), PE/PULSE OUT ※ Manually switch with coaxial line |
Blending Mode |
Common mode/differential mode ※ Manual switching with short circuiting plug |
Power capacity of the tested equipment (EUT) |
单相 AC240V/DC125V 16A(L(+),N(-),PE) |
power supply |
AC100~240V 50Hz/60Hz |
Temperature/humidity range for use |
15~35℃ 25~75℃ |
Size |
(W) 430 × (H) 249 × (D) 540mm (excluding protrusions)/approximately 20kg |
High voltage coaxial joint |
NMHV Noiseken |
attachment |
Coaxial line 30cm (02-00013A): 2 pieces, SG short-circuit plug (02-00106A): 1 piece, SG connection wire (05-00103A): 1 piece, output panel: 1 piece, AC power cord: 1 piece, operation manual: 1 volume, accessory package: 1 piece |
Attachment:
name |
model |
quantity |
description |
coaxial cable |
02-00013A |
2 pieces |
Set line for pulse output, 30cm |
SG sets short-circuit plug |
02-00106A |
1 piece |
NMHV NOISEKEN specification plug |
SG connection line |
05-00103A |
1 piece |
Grounding connection wire |
Output panel |
|
1 piece |
NMHV NOISEKEN specification panel, installed |
power cord |
|
1 piece |
Single phase terminal lifting type AC240V/DC60V 16AMax |
Operation Manual |
|
1 volume |
|
attachment package |
|
1 piece |
|
High frequency noise simulatorRecommended configuration:
High frequency testing standards |
Model Configuration |
description |
Japanese JIS Industrial Standards |
INS-S220 |
High frequency noise simulator; |
EFTD |
Pulse group experimental table; |
Other testing options:
Non standard options |
IJ-4050 |
Three phase coupling network: AC415V 50A |
18-00084A |
Output panel; Provide plugs for plugging in |
00-00017A |
Attenuator for waveform observation; DC~2GHz:40dB |
00-00011A |
Attenuator; 20dB • N-type connector |
15-00014A |
Coupling clip; Capacitive coupling, wire diameter 20mm, high input voltage: 4000V |
CA-805B |
Coupling clip (capacitive coupling); Device for testing signal lines |
CA-806 |
Coupling clip; Magnetic field coupling, wire diameter 27mm, high input voltage: 2000V |
H2-B |
EMS probe kit; Accessories for radiation testing |