The touch screen beating degree tester (professional model) is an instrument developed by our company in accordance with national standards for detecting the degree of pulp beating, also known as tapping degree. The instrument is mainly used to measure the filtration rate of diluted pulp suspension and is a routine instrument in the laboratory of pulp and paper enterprises.

Touch screen beating degree tester (professional model)It is an instrument developed by our company in accordance with national standards for detecting the degree of pulp beating, also known as tapping degree. The instrument is mainly used to measure the filtration rate of diluted pulp suspension and is a routine instrument in the laboratory of pulp and paper enterprises. By using excellent technology and high-quality materials for manufacturing, it can provide higher measurement accuracy and more stable performance. By using the Schober Rigler method principle, the beating degree of pulp can be accurately measured, and the results can be expressed through SR numerical values, with stable measurement performance.
Standard Basis
|
GB/T 3332-1982 Determination of beating degree for pulps (Schopper Rieger method)
ISO5267/1
|
Technical Parameters
Ø Measurement range: (0~100) SR, (5~95) SR effective
Ø Measurement accuracy: 0.1g
Ø Analysis accuracy: 0.1SR
Ø Display mode: 5-inch color touch screen, integrated operation control and test result display
Ø Measurement data: Display the beating degree value, real-time slurry, testing time, and testing time setting
Ø Operation mode: Automatic release test, desktop style
Ø Result display: Automatically read test results
Ø Sealing adjustment method: automatic locking sealing
Ø Side tube adjustment: With a side tube adjustment knob, the remaining volume can be adjusted freely
Ø Overflow outlet drainage time: 149 ± 1s
Ø Remaining volume: (7.5~8.0) ml
Ø Cone rising speed: 100 ± 10mm/s
Overflow calibration configuration: equipped with overflow calibration adjustment system
Ø Overflow calibration: Comes with internal overflow calibration configuration
Ø Beating value calibration: Beating value calibration can be performed
Ø Slurry value calibration: Slurry value calibration can be performed
Ø Power supply: 220V, 50Hz
Ø Dimensions: 410 * 360 * 760mm
Ø Weight: 25kg