Which is a good carbon flux analyzer

NegotiableUpdate on 05/05
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin
Overview
The backpack style trace gas analyzer (carbon flux) adopts a durable and portable design, which can provide high-performance measurement of parameters such as CO2, CH4, and H2O; Using cavity enhanced non dispersive infrared technology (infrared wavelength filtering GFC) to achieve high-precision measurement of greenhouse gases, with an accuracy of up to ppb level, supporting online viewing and analysis of measurement results; It has the characteristics of continuous measurement of three components, field deployment, and long-term stable operation. It can also transmit data through the network, making remote control easy and meeting scientific research needs to the greatest extent possible.
Product Details

Product Overview

Backpack type trace gas analyzer (carbon flux)Adopting a durable and portable design, it can provide high performanceCO2TheCH4andH2Measurement of parameters such as O; Using cavity enhanced non dispersive infrared technology (infrared wavelength filtering GFC) to achieve high-precision measurement of greenhouse gases, with an accuracy of up to ppb level, supporting online viewing and analysis of measurement results; It has the characteristics of continuous measurement of three components, field deployment, and long-term stable operation. It can also transmit data through the network and easily achieve remote control,biggerLimit to meet scientific research needs.

System Features

Backpack type trace gas analyzer (carbon flux)Adopting infrared wavelength filtering technology(GFC and Long Path Gas Absorption Cell (L-Cell) technology can detect ultra-low range gas concentrations.

Customize ultra narrow band filters for target gas absorption bands to minimize cross interference in multi-component gas measurements.

CO2TheCH4The sensor uses infrared wavelength filtering detection technology,CO2The use of dual beam detection technology can effectively solve cross interference.

Modular design: The light source, sensor, core circuit, gas chamber, etc. adopt modular design, which has high reliability, good scalability, and easy maintenance.

Adopting a multi-level constant temperature control system to ensure that all components inside the movement are at a constant working environment temperature, thereby reducing system drift.

Backpack style design, more portable.

Technical Specifications

measurement technology

GFC+L-Cell (cavity enhanced non dispersive infrared spectroscopy technique)

Measurement rate

1S/times

traffic

0.8L/min±0.5L/min

Measuring gases

CO2

CH4

H2O

measurement range

0-10000ppm

0-100 ppm

0-60000ppm

precision(1σ)

4ppm@400ppm , 2 seconds; 2ppm@400ppm , 10 seconds

20ppb@2 ppm, 2 seconds; 10ppb@2 ppm, 10 seconds

50ppm@10000ppm , 2 seconds; 25ppm@10000ppm , 10 seconds

Response time(T10-T90)

/

from 0-2 PPM ≤ 10 seconds

/

Total weight

14kg (host)

size

568mm in length, 300mm in width, and 261mm in height