Boiler water softening water hardness tester
What is softened water? Why monitor the hardness of water quality?
Water can be classified into soft and hard types, with the main criterion being its hardness, which refers to the concentration of calcium and magnesium ions in the water. Softened water refers to the process of dissolving water into water
Hardness (mainly referring to the removal or reduction of calcium and magnesium ions in water) to a certain extent. During the softening process of water, only the hardness decreases, while the total content
The salt content remains unchanged.
The detection and analysis of water hardness is an important task in water quality analysis, which affects public production and life safety. It is an indicator of water quality
Important indicators.The hardness of water is the main factor that forms scale and affects product quality. Therefore, monitoring the total hardness of water refers to the total amount of calcium and magnesium in the water
The determination,It can provide a basis for determining water quality and conducting water treatment.

Boiler water softening water hardness testerAqualysis800 product features:
Data storage, fault recording, RS485, 4-20 mA output
❏ Automatic calibration, automatic diagnosis, and monitoring
Compact design, 400x300x180mm/4Kg
Low maintenance and reagent consumption
❏ LCD color screen display in both Chinese and English
Continuous measurement or interval measurement (0~360min)
Four programmable contact outputs
❏ 2-channel programmable digital input (can be linked with external devices)
Easy to install and simple to operate
On site case display pictures:


Boiler water softening water hardness testerTechnical Specifications:
Specification and model: Aqualysis800
Measurement method: Titration colorimetric method
Measurement time: about 3 minutes, depending on the hardness of the water
Accuracy: ± 2% of the upper limit of the selected reagent
Repeatability: ± 1% of the upper limit of the selected reagent
Power supply: 85-265 VAC, 47-63Hz, power consumption: 25VA (during operation), 3.5VA (during standby)
Protection level: IP65
Size: 400x300x180mm (WxHxD)
Display: LCD color screen display
Optional units: dH, ° f, ppmCaCO3, mmol/l
Output: 1. 4-channel programmable relay output (max.250V, 4A)
2. 1 channel 4-20 mA signal, max 750 Ω
3. 1 RS485 output
Input: 1. IN1 input (start analysis/flow control switch/water meter)
2. IN2 input (reset device)
Analysis cycle: time interval measurement (0~360min)/external signal/flow signal
Rinse time: adjustable (15-1800s)
Water quality requirements: pH: 4-10; Iron:<3 ppm; Copper:<0.2 ppm=''>
Temperature: Environment: 10 ° C-40 ° C, Water sample: 5 ° C-40 ° C
Aqualysis800Water hardness analyzerReagent type and measurement range:
(The measurement range depends on the selected reagent, and the measurement error is about ± 2% of the measurement value. 500ml, valid for 2 years)
parameter |
reagent |
0dH |
0f |
ppm CaCO3 |
mmol/l |
Water hardnessreagent |
TH7001 |
0.03-0.35 |
0.045-0.626 |
0.447-6.258 |
0.004-0.063 |
Water hardnessreagent |
TH7003 |
0.07-1.04 |
0.134-1.878 |
1.341-18.78 |
0.013-0.188 |
Water hardnessreagent |
TH7010 |
0.25-3.48 |
0.447-6.258 |
4.470-62.58 |
0.045-0.626 |
Water hardnessreagent |
TH7030 |
0.74-10.43 |
1.341-18.78 |
13.41-187.8 |
0.134-1.878 |
Water hardnessreagent |
TH7050 |
1.24-17.38 |
2.235-31.29 |
22.35-312.9 |
0.223-3.129 |
Water hardnessreagent |
TH7100 |
2.48-34.76 |
4.470-62.58 |
44.70-625.8 |
0.447-6.258 |