Online residual chlorine detector for secondary water supply

NegotiableUpdate on 04/24
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin
Overview
The PM8202CL online residual chlorine analyzer for drinking water and sewage can be paired with a Bsens650 sensor and a BAF615 flow channel to monitor residual chlorine, chlorine dioxide, and ozone in water. It uses a constant voltage principle, stable measurement, simple installation, and good cost-effectiveness. It is typically used in drinking water, hospital wastewater, secondary water supply, industrial process water, etc.
Product Details

Online residual chlorine analyzer for drinking water and sewage


The PM8202CL residual chlorine/chlorine dioxide/ozone online analyzer uses the constant voltage principle to measure residual chlorine/chlorine dioxide/ozone in water.

This method utilizes a stable potential applied between the polarizing electrode and the reference electrode, and different measured components generate different current intensities at this potential.

The instrument calculates the concentration of the measured component by collecting and analyzing the current signal.


Online residual chlorine analyzer for drinking water and sewagePM8202CL product features:


No need for pre-treatment of water samples;

No need for any colorimetric reagents;

No need to replace the membrane and electrolyte;

Anti interference dual platinum electrode, reliable measurement;

Constant flow channel maintains stable flow rate;

One machine for three purposes, capable of measuring residual chlorine, chlorine dioxide, and ozone;

● Chinese and English interface, easy to operate;

Password protection to prevent human error;

● 4-20 mA.RS485Modbus;

● One multi-functional relay output;

● Wall mounted, panel mounted, and pipe mounted installations.

余氯.jpg

Online residual chlorine analyzer for drinking water and sewagePM8202CL Technical Parameters:



measurement range

0.00~20.00ppm,0.00-200.00ppb

resolution

0.001ppm,0.01ppm

precision

±2%F.S

temperature compensation

-10~130 ℃ manual/automatic; (NTC10K/PT1000)

temperature

Working temperature: -10~70.0 ℃; Storage temperature: -20~70.0 ℃

display

Large backlit dot matrix LCD

language

Chinese/English

storage

600000 pieces of data

power supply

90-260VAC,50/60Hz; 24VDC optional

transmitter output

2-channel isolated transmission 4-20mA output, maximum loop 500 Ω, 0.1% F.S, adjustable measurement value and temperature

communication function

RS485 Modbus

Cleaning output

Cleaning interval: adjustable from 0.1-1000h, cleaning time: adjustable from 1-1000s

alarm output

2 independent Hi/Lo alarm points with hysteresis settings, 5A/250VAC/30VDC

Installation method

Wall mounted, pipe mounted, panel mounted

Protection level

IP65

Overall Dimensions

144×144×108mm

cutout size

138×138mm

weight

0.87KG


Electrode parameters:

model

Bsens650

measurement range

0~2.000,0~20.00ppm

resolution

0.001/0.01ppm

precision

±2%F.S.

electrode material

Glass/Double Platinum Ring

Operating Temperature

0~70.0℃

Z high withstand voltage

5bar

connection method

fixed

thread size

PG13.5

cable length

3m

application

Measurement of residual chlorine/chlorine dioxide/ozone values in water