The following are the volume and weight specifications of the Fuli Instrument AA1700 atomic absorption spectrophotometer, compiled from comprehensive and reliable information:
Volume parameters
Host size
610 mm (length) x 500 mm (width) x 450 mm (height)
Graphite furnace module size (if configured)
240 mm (length) x 500 mm (width) x 420 mm (height)
Installation suggestions
Reserve an operating space of ≥ 1 m ² (including auxiliary equipment such as gas cylinders and workstations)
Weight parameters
Component weight remarks
Host 45 kg including basic optical and electronic systems
60 kg graphite furnace module requires independent power supply and heat dissipation support
The transportation weight of the whole machine is ≥ 105 kg, including packaging and accessories (it is recommended to handle the logistics layer by layer)
Key precautions
Power supply and power consumption
Host power consumption: 150 W; Graphite furnace module power consumption: 3600 W (requires independent circuit)
Heat dissipation requirements
Reserve a ventilation distance of ≥ 20 cm on the side of the equipment to avoid high temperature operation affecting stability
Security protection
Configure safety mechanisms such as acetylene leakage alarm and undervoltage protection
Supplementary explanation
The host adopts a compact design (reducing the volume by about 15% compared to similar devices);
The graphite furnace module needs to be strictly grounded to prevent high-frequency interference.
Suggest contacting the supplier to obtain detailed transportation packaging drawings and laboratory deployment plans.
The following is a comprehensive summary of the core specification parameters of Fuli Instrument AA1700 atomic absorption spectrophotometer:
Basic parameters
Host size
610 mm (length) x 500 mm (width) x 450 mm (height)
Graphite furnace module size (optional)
240 mm (length) x 500 mm (width) x 420 mm (height)
Weight
Host: 45 kg; Graphite furnace module: 60 kg (requires independent power supply)
Performance indicators
Parameter specifications
Optical system dual beam design, flat grating monochrome element
Detector photomultiplier tube (PMT)
Resolution Mn 279.5nm/279.8nm, valley ratio < 15%
Baseline stability: Cu element<0.003 μ g/ml
Repeatability (RSD) ≤ 0.5%
Detection limit: Cu element ≤ 0.003 mg/L
Atomization system
火焰原子化器
Support automatic switching between acetylene and laughing gas (high temperature element detection)
Automatic ignition/shutdown, automatic flow adjustment
Graphite Furnace Atomizer (optional)
Power consumption: 3600 W (requires dedicated circuit)
Temperature control accuracy: ± 0.2 ℃ (optional precision temperature control module)
Security and Automation
Protection mechanism
Acetylene leakage alarm, undervoltage protection, abnormal power outage protection
Intelligent functions
Fault self inspection, automatic background correction, real-time monitoring of gas path status
Atomizer
Standard metal sleeve glass high-efficiency atomizer (compatible with mainstream domestic and foreign brands)
Auxiliary configuration
Power requirements
Host: AC 220V/50Hz, power consumption 150W
Graphite furnace: AC 220V/50Hz, power consumption 3600W
Space requirements
Operating area ≥ 1 m ² (including gas cylinder and workstation)
Reserve a heat dissipation distance of ≥ 20 cm on the side
Software and extensions
Workstation: Intelligent expert application software, supporting fault diagnosis and data management
Structural design: Compact body (reduced by about 15% in volume compared to similar models), suspended optical path for improved stability
Note: The graphite furnace module needs to be strictly grounded to prevent high-frequency interference; Suggest obtaining installation drawings and operation manuals for specific configurations from suppliers.