Based on the physical parameter information of Beijing Puxi TAS-990 series atomic absorption spectrophotometer in the search results, the volume and weight specifications are summarized as follows:
Physical specification parameters
Project specifications
The total weight of the machine is about 35kg (including standard configuration)
The size of the host is not clearly indicated in the search results (designed as a compact desktop instrument)
Supplementary explanation
Installation space requirements
Reserve a laboratory bench area of ≥ 0.8 m ² (including supporting operating space);
Reserve a heat dissipation space in the rear (recommended ≥ 20cm) to ensure stable operation of the instrument.
Extended accessory impact
If equipped with accessories such as an automatic sampler (such as TAS-990G), graphite furnace, or atomizer, it may result in additional weight and space occupation.
For example:
The weight of the dedicated graphite tube is about 500g per piece;
The weight of the metal sleeve glass atomizer is not disclosed (needs to be installed separately).
Note: For specific host dimensions, it is recommended to directly refer to the equipment operation manual or contact the supplier for accurate data.
The following is a comprehensive summary of the core specification parameters of Beijing Puxi TAS-990 series atomic absorption spectrophotometer:
Optical system parameters
Project specifications
Wavelength range 190-900nm
Spectral bandwidth automatically switches between five levels: 0.1nm, 0.2nm, 0.4nm, 1.0nm, and 2.0nm
Wavelength accuracy ± 0.25nm (standard model)/± 0.15nm (SUPER model)
Wavelength repeatability ≤ 0.15nm (standard model)/≤ 0.05nm (SUPER model)
Baseline drift ≤ 0.004Abs/30min (static)
Background correction
Deuterium lamp calibration: capable of calibrating 1.0Abs background
Self suction correction: capable of correcting 1.8Abs background (SUPER models up to 1.8A)|
Analyze performance parameters
Flame analysis method
Project specifications
Characteristic concentration (Cu) ≤ 0.04 μ g/mL/1% (standard model)/≤ 0.02 μ g/mL/1% (SUPER model)
Detection limit (Cu) ≤ 0.006 μ g/mL (standard model)/≤ 0.004 μ g/mL (SUPER model)
Precision (RSD) ≤ 1.0%
Graphite furnace analysis method
Project specifications
Characteristic quantity (Cd) ≤ 1.0 × 10 ⁻¹² g (standard model)/≤ 0.4 × 10 ⁻¹² g (SUPER model)
Detection limit (Cd) ≤ 1.0pg (standard model)/≤ 0.5pg (SUPER model)
Precision (RSD) ≤ 3.0%
Hardware configuration and functionality
Project specifications
Optical system dual beam design, aberration correcting C-T monochromator
Elemental lighting system with eight automatic turret lights, supporting automatic positioning and fine-tuning
Burner type: Pre mixed all titanium combustion head (flame method)
Horizontal heating design of graphite furnace technology, optimization of temperature uniformity
Security protection
Real time flame monitoring+automatic cutting off of acetylene leakage
Cooling water flow monitoring+abnormal pressure protection|
|Software Platform | AAWin 2.0 Workstation, supporting standard curve method, standard addition method, GLP specification|
Physical parameters
Project specifications
The total weight of the machine is about 35kg (including standard configuration)
Installation requirements
Experimental bench area ≥ 0.8m ² (including operating space)
Rear heat dissipation space ≥ 20cm|
Key Explanation
Model difference
Suffix identification:
F: Flame type
G: Graphite furnace type
FG: Flame+Graphite Furnace Manual Type
AFG: Flame+Graphite Furnace Automatic Type
SUPER model: higher accuracy (such as wavelength repeatability ≤ 0.05nm)
Extended functionality
Automatic sampler (such as TAS-990G) supports 60 sample grades and 8 standard sample positions
The gas flow rate adopts a mass flowmeter, which has higher accuracy than a rotor flowmeter
Operational requirements
The flame method requires an air compressor pressure of 0.2-0.3MPa and an acetylene pressure of 0.05-0.06MPa
Cooling water needs to be supplied before graphite furnace analysis, and the flow rate should be started after reaching the standard
Note: Specific parameters are subject to the equipment operation manual. Additional accessories such as atomizers and automatic samplers may require additional weight and space.