For performance testing purposes ECG simulator, which can be used standalone or connected to a PC, outputs high-precision and low-noise analog signals, providing various noise, breathing, and lead dropout simulation signals, suitable for research and compliance testing,Meets international medical standard testing requirements:IEC 60601-2-25、IEC 60601-2-27、IEC 60601-2-47、GB 9706.225、GB 9706.227、YY 9706.247、YY1079、YY1139、YY0782、YY0885、JJG 760、JJG 1042、ANSI/AAMI EC11、ANSI/AAMI EC13
·Built in compliance Electronic circuits required by ECG standards
·Adjustable testing parameters and options provide testing flexibility
·NST noise simulation-Electrode Motion Artifact / Muscle Noise / Baseline Wander
·Multi style signal simulation -Impedance breathing, modulation breathing (baseline/amplitude/frequency), lead dropout signal
·The Auto Sequence function assists users in easily developing semi-automatic testing processes
·Can load and play recorded or programmed digital waveform signals, quickly validate signal processing systems and algorithms
·Provide software development toolkits(Software Development Kit), Users can develop customized or fully automated testing software on their own
·Optional SECG Assistant (ECG Medical Standard Testing Assistance Software) simplifies testing steps and allows users to complete standard testing projects by clicking on options
·simulation Perform SpO on ECG and PPG signals2Testing with PWTT (Pulse Wave Transit Time)
·Flexible adjustment SpO2Parameters such as AC, DC, and PI values of R/IR are used to confirm the accuracy of the test substance
·Customize or load SpO2R-curve, quickly verify product performance, and assist users in confirming competitor R-curves
·built-in 14 types of ambient light signals and advanced respiratory simulation parameters for overlay testing to better simulate real-life scenarios
·synchronization ECG and PPG signals, with adjustable PWTT time difference, to test the accuracy of non-invasive blood pressure measurement algorithms





parameter |
Specifications |
Main output voltage accuracy |
± 0.5%, amplitude of 0.5mVpp or higher |
Main output voltage resolution(DAC resolution) |
0.63µV |
frequency /Pulse repetition rate accuracy |
±0.5% |
Pulse duration /Time accuracy (excluding pacing) |
±0.5ms |
Accuracy of pacing pulse width |
±5μs |
Accuracy and range of pacing pulse amplitude |
± 2mV pulse: ± 0.3% |
Pacing pulse characteristics |
rise /The descent time is 5 μ s |
Pacing pulse overshoot |
methodA, According to IEC 60601-2-27 |
Resistance tolerance |
±1% |
Capacitance tolerance |
±5% |
precision 250:1 voltage divider |
±0.1% |
sampling rate |
10kHz ± 0.05% |
DC offset |
300mV ± 1% |
DC offset |
± 1000mV ± 1% or ± 3mV |
environment |
Indoor use |
parameter |
set the scope |
default value |
Minimum adjustable scale |
|
Sine |
Frequency(Hz) |
0.05 ~ 5.00Hz |
1Hz |
0.01 |
Frequency(BPM) |
3 ~ 30000BPM |
60BPM |
1 |
|
amplitude |
(-40) ~ 40mV |
1mV |
0.01 |
|
Triangle |
Frequency(Hz) |
0.05 ~ 5.00Hz |
1Hz |
0.01 |
Frequency(BPM) |
3 ~ 30000BPM |
60BPM |
1 |
|
amplitude |
(-40) ~ 40mV |
1mV |
0.01 |
|
Square |
Frequency(Hz) |
0.05 ~ 5.00Hz |
1Hz |
0.01 |
Frequency(BPM) |
3 ~ 30000BPM |
60BPM |
1 |
|
amplitude |
(-40) ~ 40mV |
1mV |
0.01 |
|
Rectangle Pulse |
Frequency(Hz) |
0.05 ~ 5.00Hz |
1Hz |
0.01 |
Frequency(BPM) |
3 ~ 30000BPM |
60BPM |
1 |
|
amplitude |
(-20) ~ 20mV |
1mV |
0.01 |
|
pulse width |
1 ~ 300ms |
100ms |
1 |
|
Triangle Pulse |
Frequency(Hz) |
0.05 ~ 5.00Hz |
1Hz |
0.01 |
Frequency(BPM) |
3 ~ 30000BPM |
60BPM |
1 |
|
amplitude |
(-20) ~ 20mV |
1mV |
0.01 |
|
pulse width |
1 ~ 300ms |
100ms |
1 |
|
Exponential |
Frequency(Hz) |
0.05 ~ 5.00Hz |
1Hz |
0.01 |
Frequency(BPM) |
3 ~ 30000BPM |
60BPM |
1 |
|
amplitude |
(-20) ~ 20mV |
1mV |
0.01 |
|
ECG Waveform |
P-wave amplitude |
0 ~ 2mV |
0.2mV |
0.01 |
P-wave interval |
50 ~ 100ms |
100ms |
1 |
|
QRS amplitude |
(-20) ~ 20mV |
1mV |
0.1 |
|
QRS interval |
5 ~ 200ms |
100ms |
1 |
|
PR interval |
0 ~ 200ms |
160ms |
1 |
|
QT interval |
100 ~ 900ms |
350ms |
1 |
|
ST variation value |
(-2) ~ 2mV |
0mV |
0.01 |
|
T-wave amplitude |
0 ~ 5mV |
0.2mV |
0.01 |
|
Frequency(Hz) |
0.02 ~ 6Hz |
1Hz |
0.01 |
|
Frequency(BPM) |
1 ~ 360BPM |
60BPM |
1 |
|
amplitude |
(-20) ~ 20mV |
1mV |
0.1 |
|
parameter |
set the scope |
default value |
Minimum adjustable scale |
|
DC offset |
(-1000) ~ 1000mV |
0mV |
1 |
|
Pacing signal |
amplitude |
(-700) ~ 700mV |
0mV |
1 |
interphase |
0.1 ~ 3ms |
2ms |
0.1 |
|
Pacing rate |
10 ~ 360BPM / Synchronization |
10BPM |
1 |
|
Overshoot amplitude |
0 ~ 5mV / Amplitude automatic |
0mV |
0.01 |
|
Overshoot time constant |
0 ~ 100ms |
0ms |
1 |
|
Baseline overload test |
frequency |
50 / 60Hz |
50Hz |
- |
620k Ω/4.7 nF (open as short circuit) |
on / off |
Off |
- |
|
Overlay sine wave |
on / off |
Off |
- |
|
noise |
Power frequency noise |
50Hz / 60Hz / White |
50Hz |
- |
amplitude |
0.01 ~ 5mV |
0.1mV |
0.01 |
|
parameter |
set the scope |
default value |
Minimum adjustable scale |
|
Play source files |
sampling rate |
upper limit 20K Hz |
- |
- |
Noise loading synthesis |
Types of noise |
EM - Action Noise |
EM |
- |
magnification |
0.1 ~ 10 |
1 |
0.1 |
|
Automatic measurement items |
heart rate |
30, 60, 120, 180, 200, 250 |
- |
- |
QRS 波振幅 |
0.5, 1, 2, 5 |
- |
- |
|
QRS interval |
40, 70, 80, 100, 120 |
- |
- |
|
Automatic measurement items |
Amplitude(mV) |
± 2, ± 100, ± 300, ± 500, ± 700 |
- |
- |
Interval(ms) |
0.1, 0.5, 1, 2 |
- |
- |
|
synchronization |
synchronization /Not synchronized, set at 80BPM |
- |
- |
|
pulse count |
Single pulse /Double pulse - lead 150ms/Double pulse - lead 250ms |
- |
- |
|
Overshoot time(ms) |
0, 4, 10, 20, 50, 100 |
- |
- |
|
QRS/T |
on/off |
- |
- |
|
Frequency scan (sine wave) |
starting frequency |
0.67 ~ 500Hz |
0.67Hz |
0.01 |
End frequency |
0.05 ~ 500Hz |
150Hz |
0.01 |
|
interphase |
10 ~ 180s |
30s |
0.01 |
|
Frequency scanning(ECG) |
BPM |
- |
3 ~ 30BPM |
- |
interphase |
- |
30s |
- |
|
Lead detachment detection |
Lead detachment detection |
short circuit /10M Ω/20M Ω/open circuit |
short circuit |
- |
parameter |
Specifications |
|
heart rate |
set the scope |
10 ~ 300BPM |
Minimum adjustable scale |
1BPM |
|
accuracy |
±1BPM |
|
LED DC signal output |
set the scope1 |
100 ~ 3000mV |
Minimum adjustable scale |
1mV |
|
LED AC signal output |
set the scope1 |
0.75 ~ 30mV |
Minimum adjustable scale |
0.01mV |
|
PI perfusion index (AC value/DC value)2 |
set the scope |
0.025%~30% (AC/DC, varying with AC or DC value) |
accuracy |
NA |
|
LED scanning rate |
rate |
10K (raw data mode) |
accuracy |
±5μs |
|
PD sampling rate |
rate |
250K (single channel) |
accuracy |
±5μs |
|
PD response time |
rise time |
1 μ s typical value3 |
fall time |
1 μ s typical value3 |
|
SpO2% |
set the scope |
1% ~ 100% |
Minimum adjustable scale |
1% |
|
accuracy |
91%~100%: ± 1%+accuracy of the test substance |
|
MCX LED voltage |
amplitude |
Communication level x 100 |
accuracy |
±5% |
|
Note: | ||
parameter |
Specifications |
|
Time difference(PTTp, PTTf) setting range |
0~1999ms (decreases with increasing pulse rate per minute) |
|
Time difference(PTTp, PTTf) Minimum Adjustable Scale |
1ms |
|
Time difference(PTTp, PTTf) accuracy |
±1ms |
|
ECG parameter specifications for PWTT testing mode | ||
heart rate |
set the scope |
10 ~ 300BPM |
Minimum adjustable scale |
1BPM |
|
ECG amplitude |
set the scope |
0 ~ 20mV |
Minimum adjustable scale |
0.1mV |
|
T-wave amplitude |
set the scope |
0 ~ 5mV |
Minimum adjustable scale |
0.01mV |
|
QRS interval |
set the scope |
0 ~ 200ms |
Minimum adjustable scale |
1ms |
|
SpO of PWTT test mode2Parameter Specifications | ||
heart rate |
set the scope |
10 ~ 300BPM |
Minimum adjustable scale |
1BPM |
|
PI perfusion index |
set the scope |
0.025% ~ 30% |
Minimum adjustable scale |
0.001% |
|
LED DC signal output |
set the scope |
100 ~ 3000mV |
Minimum adjustable scale |
1mV |
|
LED AC signal output |
set the scope |
0.75 ~ 30mV |
Minimum adjustable scale |
0.01mV |
|
Product Part Number |
Product Description |
100-EC00002 |
SECG 5.0 AIO Multi physiological Signal Simulator, supports single machine and PC software operation, can output 12 lead (RA, LA, LL, N, V1-V6) analog signals. |
Product Part Number |
Product Description |
100-AE00007 |
PPG-2TF-660 penetrating blood oxygen performance testing module supports SECG 5.0 AIO for PWTT/SpO2Test mode. |
K29-0300601 |
MCX (RF) male to BNC cable x 1 |
Product Part Number |
Product Description |
HA0-SE0G001 |
Medical standards IEC 60601-2-25:2011 Performance Testing Auxiliary Software |
HA0-SE0G002 |
Medical standards IEC 60601-2-27:2011 Performance Testing Auxiliary Software |
HA0-SE0G003 |
Medical standards IEC 60601-2-47:2012 Performance Testing Auxiliary Software |
HA0-SE0G004 |
Medical standards YY0782-2010 Performance Testing Auxiliary Software |
HA0-SE0G005 |
Medical standards YY0885-2013 Performance Testing Auxiliary Software |
HA0-SE0G006 |
Medical standards YY1079-2008 Performance Testing Auxiliary Software |
HA0-SE0G007 |
Medical standards YY1139-2013 Performance Testing Auxiliary Software |
HA0-SE0G008 |
Medical standards GB 9706.225-2021 Performance Testing Auxiliary Software |
HA0-SE0G009 |
Medical standards GB 9706.227-2021 Performance Testing Auxiliary Software |
HA0-SE0G010 |
Medical standards YY 9706.247-2021 Performance Testing Auxiliary Software |