Quotation for Density Meter Calibration Device

NegotiableUpdate on 01/14
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin

Overview

The density meter calibration device complies with the requirements of JJG42-2011 "Verification Regulations for Working Glass Floats" and is suitable for the calibration, subsequent calibration, and in-service calibration of fixed quality working glass floats (hereinafter referred to as floats) such as density meters, petroleum density meters, alcohol meters, sugar meters, milk meters, soil meters, etc.

Product Details

密度计检定装置

Density meter calibration device

Overview

HK-42A Density meter calibration device(Static weighing method), in compliance withThe requirements of the JJG 42 Verification Regulations for Liquid Static Weighing Method are applicable to the verification, subsequent verification, and in-service verification of fixed mass working glass floats such as density meters, petroleum density meters, alcohol meters, sugar meters, milk meters, soil meters, etc.

Features

1.High precision cooling constant temperature water bath combined with customized double-layer thick glass calibration cylinder, with a constant temperature field

2.The instrument's large touch screen is equipped with an intelligent calculation program, which can quickly and efficiently obtain calibration correction values and can be stored,USB table export

3.Large torque lifting mechanism with remote control button for two lifting modes, convenient to use

4.Integrated design, high-definition camera, clear alignment with density meter markings

5.Equipped with imported scales and high-precision temperature, humidity, and atmospheric pressure sensors to ensure accurate data collection

6.Equipped with density weights and density measurement programs, convenient for measuring liquid density

Technical Specifications

1.electricity Source:AC 220V±10% 50Hz

2.Heating power:≤2000W

3.Accuracy of external constant temperature water bath:±0.01℃

4.Inner diameter of glass calibration cylinder: diameter110mm, 500mm high

5.Temperature accuracy of bath collection:0.001℃

6.Room temperature acquisition accuracy:5℃~50℃,±0.5℃

7.Humidity collection accuracy:0~85%RH,±3%RH(20℃)

8.Atmospheric pressure collection accuracy:±0.3hPa

9.sky Ping: Resolution of Sartorius electronic balance0.1mg

10.Environmental requirements: Temperature18℃~22℃; Humidity ≤ 85% RH