Production of Karl Fischer Moisture Analyzer

NegotiableUpdate on 01/16
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin

Overview

The V-310S Karl Fischer moisture analyzer adopts a 5.7-inch capacitor high-definition color touch screen, which is effortless and easy to operate like a mobile phone. The operation interface is simple and intuitive, and the navigation based operation method reduces learning costs and makes it faster to get started. *The circuit and program design have advantages such as high precision, fast electrolysis speed, short balancing time, accurate blank deduction, and reliable analysis results.

Product Details

Automatic Karl Fischer moisture analyzerMainly used for testing various chemical products, high-tech materials, pharmaceuticals, and petroleum products, especially for samples with low moisture content, with high accuracy and good repeatability, such as lithium battery electrolysis, which can detect below 10ppm and maintain good repeatability.

Not affected by the environment, automatic drift monitoring

The system automatically detects environmental drift and can accurately deduct it, ensuring that the moisture detection results are not affected by the environment and mixing speed;

Connecting the Karl Fischer oven can also automatically detect and deduct moisture in the carrier gas, further reducing the requirements for the carrier gas and lowering user usage costs;

Multiple endpoint judgment methods, suitable for various reagents

Built in multiple endpoint judgment methods: relative drift; Relative drift+time; Measure potential+relative drift; Multiple endpoint judgment methods such as measuring potential and time allow users to choose the appropriate endpoint judgment method based on the environment and sample conditions.

Using dynamic drift algorithm to determine the endpoint, it is not affected by the environment and can be used with confidence even in high temperature and high humidity areas;

The endpoint potential can be set, suitable for different manufacturers and models of Karl Fischer reagents.

The system has multiple built-in methods, and the result units can be customized

The system has 5 built-in methods and 8 formulas, suitable for various solid, liquid, gas and other samples;

Automatic calculation of measurement results, including percentages, ppm, mg, etc; GLP compliant experimental records, including measurement time, moisture content, percentage, system time, operator, etc;

Carefully select various industrial grade imported components to ensure the reliability of the instrument

Automatic Karl Fischer moisture analyzerAll internal components are imported industrial grade components from major manufacturers to ensure the stability and reliability of the instrument

Capacitive touch screen makes navigation menu operation more convenient

The instrument adopts a capacitive touch screen, which is different from traditional resistive touch screens, with more sensitive touch and simpler operation;

Navigation style menu, clear function options, easy to use instrument.

Display more comprehensive and intuitive content

Simultaneously displaying water content, measurement time, PPM, percentage content, electrolysis current, potential, and system time online, the experimental process is clear at a glance and problems can be detected in a timely manner;

Electrolytic current stepless regulation, test results can be calibrated

The electrolysis current is 0-500mA, which can be customized and adjusted up to 500mA, and the electrolysis rate can reach 2.8mg/min.

The test results can be automatically calibrated by filling in the correction factor based on the calibration results.

Automatic Karl Fischer moisture analyzerTechnical Specifications

name

Technical Specifications

model

V310-S

measurement range

1Mg ---300mgwater

Measuring speed

2.8mg H2O/min()

measurement accuracy

3Mg (10ug--100Mg H2O) ;≤ 0.3% (water content>100Mg H2O

resolution

0.1Mg H2O

Relative moisture content range

1ppm-99%Display accuracy 0.001ppm

Electrolytic current

0-500mAStepless adjustment

electrolytic electrode

Platinum belt with diaphragm

Featured Features

5There are 8 calculation formulas to choose from, with sample units of ppm,%, mg/L, and mg/kg that can be freely switched

injection delay

0-9999customize

electrolysis time

0-9999Custom (0 means unlimited electrolysis time)

data storage

500article

display screen

5.7Inch capacitive touch screen

Calibration function

yes

printer

Comes with a 58mm wide line thermal printer

Usage environment

Temperature+5~+40 ℃; Humidity<80%<>

power supply

110-250V 50Hz

instrument size

495x330x325mm

Instrument weight

6kg