Welcome Customer !

Membership

Help

Shanghai Qiantuo Trading Co., Ltd
Custom manufacturer

Main Products:

plast-mach>Products

Operating Standard FMU30-AAHEAAGGF German E+H Ultrasonic Level Gauge

NegotiableUpdate on 01/17
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin

Overview

Operating standard FMU30-AAHEAAGGF German E+H ultrasonic level gauge sensor directly emits ultrasonic pulse signals to the surface of the medium

Product Details

Operating Standard FMU30-AAHEAAGGF German E+H Ultrasonic Level Gauge

The sensor directly emits ultrasonic pulse signals to the surface of the medium.

The pulse signal is reflected on the surface of the medium, and the reflected signal is received by the sensor.

The instrument measures the time difference between transmitting and receiving pulse signals.

Calculate the distance between the sensor diaphragm and the surface of the medium based on time difference and sound velocity.

[AA] Certification: Non explosion proof occasions

[H] Display, operation: on-site envelope display, buttons

[E] Electrical connection: Cable sealing sleeve M20, IP68

[AA] Sensor; Range; Blind spot: 1-1/2 '; 5m liquid/2m solid; 0.25m

[GGF] Process connection: Thread ISO228 G1-1/2, PP

Power/Communicationtwo-wire system

FMU30 ultrasonic level gauge liquid range: 5m OR 8m

Resolution: 1mm

精度: ±3 mm OR 0.2%

Blind spot: 25cm or 35cm

Power supply: 24VDC (standard); 220VAC (optional)

FMU30 ultrasonic level gauge display: clear multi line text graphic display, envelope line display

Output: 4-20mA

Process connection: 1-1/2 "(5m), 2" (8m)

Shell material: F16 plastic transparent display cover

Advantages and Characteristics of FMU30 Ultrasonic Level Gauge

It is an ideal upgrade substitute for FMU230 and FMU231

Good protection level and high cost-effectiveness

Menu guidance function, easy to achieve on-site operation

measurement accuracy+/-3 mm or+/-0.2% of the set range

ambient temperature-20 °C. ..+60 °C (-4 °F...+140 °F)

process temperature-20 °C. ..+60 °C (-4 °F...+140 °F)

Process pressure (absolute pressure)/overpressure limit value0.7 bar . .. 3 bar abs (10 psi ... 44 psi)

Main liquid receiving componentsPP/EPDM

process connectionG 1-1/2'

G 2'

Blind spot distance1 1/2 'sensor: 0.25m

2 'sensor: 0.35m

measure distance1-1/2 'sensor: 2 m (6.6 ft)

2 'Sensor: 3.5 m (11 ft)

output4...20 mA

Application restrictionsPay attention to the measurement range diagram

Operating Standard FMU30-AAHEAAGGF German E+H Ultrasonic Level Gauge