Compliant with standards
《HJ618-2011ambient airPM10andPM2.5The method of determining weight
How to translate《HJ656-2013Environmental air particulate matter(PM2.5)Manual monitoring method(gravimetric method)
technical specificationsHow to translate
《
HJ836-2017Determination of low concentration particulate matter in exhaust gas from fixed pollution sources by gravimetric methodHow to translate

Equipment Usage
Used in ambient air and fixed source exhaust gasesPM2.5ThePM10Weighing of low concentration particulate matter.System IntroductionHW5500typeConstant temperature and humidity weighing systemAlso known as low concentrationparticulate matterThe weighing system and filter membrane weighing system are laboratory specific equipment used for balancing and weighing low concentration particulate matter sampling filters and filter heads in fixed pollution source exhaust gas. This system can control the internal temperature of the box with high precisionwetDegree,24the systemOvercoming the impact of temperature and humidity changes in the laboratory environment on weighing results and improving weighing performance
result
The accuracy.Constant temperature and humidity weighing system
It is to place a high-precision balance in a stable constant temperature system environment, place the sample to be weighed in the system, and balance it |
After hours, perform manual weighing |
. |
HW-5500 |
> |
Equipment parameters |
power supply |
AC220V / 50Hzrated power1000Wworking environment |
10℃~40℃ |
Overall dimensions1310*700*1300mm |
(Length * Width * Height) |
) |
Operation box size |
880*600*660 |
mm |
(Length * Width * Height) |
Equipment weight |
140KG (approximately) |
Temperature control range |
15 ℃~70 ℃ (factory set to 25 ℃; adjustable as desired) |
temperature control accuracy |
±0.3℃ |
humidity range |
25% RH to 80% RH (factory set to 50% RH; adjustable as desired) |
Moisture control accuracy±3%RH
static elimination1Ion blowing
Seismic gradeLevel 3 earthquake resistance, decoupling isolation
>
Main ConfigurationManual weighing system
setThree level seismic protection system1 set1 set of humidification system
LCD (Liquid Crystal Display)
1 set of screensampleplaceFrame 1
a
1 set of ion blowing electrostatic cleaning device (built-in)Imported 1/100000 scale(
optional
)
Technical advantages
HW5500 adopts
High strength sheet metal casing, German KNF imported motor, Japanese Omron relay module, and Panasonic imported compressor.
Due to its high-precision sensors, microcomputer digital processing technology, enclosed enclosure space, and high seismic structural design, the system fundamentally improves the reliability and maintainability of the instrument.
Temperature and humidity control instrument: The screen operation is simple, and the heating power can be adjusted separately, the compressor delay start can be set, and the service life of the compressor can be protected.
Different sample holders can be placed for weighing different types of samples.Adopting a dual cold and hot air duct design, it can automatically switch between cold air ducts for more precise temperature and humidity control.
The observation window is made of customized materials, which have the advantages of transparency, insulation, and are not prone to steam frosting.The working area of the instrument is designed for separate operation, effectively avoiding the impact of refrigeration system vibration on the balance.
The heating and cooling system is independent.Humidification system: External isolation system, all stainless steel shallow surface humidification.
● Supply air circulation system: temperature resistant and low-noise fan, effectively removing static electricity from the sample
.