Online water hardness analyzer
Aqualysis800 online water hardness monitorWidely used for monitoring the alkalinity of water quality in boiler water, softener effluent, cooling circulating water, surface water, pharmaceutical injection water, etc;
Why chooseAqualysis800 online water hardness measuring instrument?
Aqualysis 800Application of titration colorimetric principle for online monitoring of water hardness Quality control of water softening process, compact structure, simple operation, and precision High, low reagent consumption, low maintenance cost, and can operate continuously for a long time.
,
Product Features:
▶ Easy installation and simple operation
▶High cost-effectiveness and high measurement accuracy
▶ Automatic calibration, automatic diagnosis, and monitoring
▶ Low maintenance and reagent consumption
▶ Chinese and English LCD color screen display
▶ Compact design, 400x300x180mm/4Kg
▶ Continuous measurement or interval measurement (0~360min)
▶ 4-channel programmable contact output
▶ 2-channel programmable digital input (can be linked with external devices)
▶ Data storage, fault recording, RS485, 4-20 mA output
Aqualysis800 water hardness testerAccording to the principle of titration colorimetry, the selectable limit values are controlled, and accurate measurement readings are provided through extinction method. Multiple functions ensure the reliability of real-time operation.
Aqualysis800 detector parameters:
| Specification Model | Aqualysis800 (OEM available) |
| measurement method | Titration colorimetric method |
| measure time | About 3 minutes, depending on the hardness of the water |
| precision | ± 2% of the upper limit of the selected reagent |
| repetitiveness | ± 1% of the upper limit of the selected reagent |
| power supply | 85 - 265 VAC, 47-63Hz, |
| power consumption | 25VA (operating), 3.5VA (standby) |
| protection level | IP65 |
| size | 400x300x180mm(WxHxD) |
| display | LCD color screen display |
| selectable unit | °dH、°f、ppmCaCO3、mmol/l |
| output | 1. 4-channel programmable relay output (max.250V, 4A) |
| 2. 1 channel 4-20 mA signal, max 750 Ω | |
| 3. 1 RS485 output | |
| input | 1. IN1 input (start analysis/flow control switch/water meter) |
| 2. IN2 input (reset device) | |
| Analysis cycle | Time interval measurement (0~360min)/external signal/flow signal |
| Rinse time | Can be set (15~1800s) |
| water quality requirements | pH:4 – 10 |
| Iron:<3 ppm | |
| Copper:<0.2 ppm=''> | |
| Aluminum:<0.1 ppm | |
| Manganese:<0.2 ppm | |
| temperature | Environment: 10 ° C -40 ° C |
| Water sample: 5 ° C-40 ° C | |
| pressure | Ca. 0.2-5 bar (max) (recommended 1-2 bar) |