German BERTHOLD low background alpha, beta surface contamination analyzer quotation

NegotiableUpdate on 01/17
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin

Overview

The German BERTHOLD low background alpha and beta surface contamination meter is a low background surface contamination meter that can simultaneously measure the alpha and beta activities in different samples.

Product Details

German BERTHOLD low background alpha, beta surface contamination analyzerLB 2046Introduction:
LB 2046 low background alpha, beta surface contamination analyzerThe activity meter can simultaneously measure the alpha and beta activities in different samples
Widely used in: nuclear facilities or reprocessing plants, nuclear power plants or chemical plants, etc
The sample on the filter in water samples, environmental samples, or air samplers can be easily measured
The detection limit mainly depends on the size of the detector, the measurement geometry, the system background, and the sample measurement time
LB 2046 low background alpha, beta surface contamination analyzerEquipped with a 60mm diameter scintillation detector
The sample is placed in a sliding drawer, and the maximum height of the sample disc can reach 8mm, with a diameter of 60mm or 100mm
The detector can simultaneously and separately measure alpha and beta activities
The detection limit of α is about 0.03Bq, the detection limit of β (Cl-36) is about 0.06Bq, and the counting time is 1 hour
The sensitive touch graphic display screen provides users with an intuitive operating interface, which can easily and quickly operate all functional menus

637860734216697592884.jpg


German BERTHOLD low background alpha, beta surface contamination analyzerLB 2046Features:
Simultaneously measure the activity of a sample with a diameter of 60mm
Closed measurement room equipped with scintillation counter
• Easy to operate, sensitive touch graphic display screen
• Interface: Serial data interface, parallel data interface, USB, Ethernet (optional)
• Service functions: Ping, background, calibration, system testing
• Data * Storage
• Stored nuclide library
An important function is automatic ramp determination, which also determines the correct operating point
Measure and store background values, deduct alpha and beta background values, and calibrate the measured values
For different nuclides, the calibration program can calibrate the probe
LB 2046 low background alpha, beta surface contamination analyzerEquipped with system testing function, it can quickly test the performance of the instrument
A wide range of nuclide libraries are available, capable of storing measurement parameters required for various nuclide and sample types
Additionally, calibration factors, measurement units, counting times, and alarm thresholds can be determined, and automatic alarms can be triggered for high-level samples
After each measurement, the measured value, sample status, and sample ID can be printed out

LB 2046 low background alpha, beta surface contamination analyzer parameters:

detector 60 mm ZnS/PMT scintillation counter
Incident window Hostaphan 铝箔, 0.4mg/cm²
efficiency Am-241: 22%, Cl-36: 41 %, C-14: 8 %
overflow α( 在β 通道, Po-210):<10%>β (在α 通道): <2.0 e-05=''>
measurement range α: 0-5000 cps
β: 0-50000cps
Alpha/Beta Measurement Room Copper sample drawer (filter with a diameter of 60/100 mm or sample disc with a height of up to 8mm)
Detection limit (1 hour) Am-241: 0,03 Bq
Cl-36: 0,06 Bq
background Alpha channel: approximately 0.02 CPS
Beta channel: approximately 0.3 CPS
display screen 320x240 pixels, backlight LED (auto turn off)
power supply 90 - 260 V AC ,50/60 Hz
weight About 45 kg
size 330 x 245 x 270 mm
Operating Temperature 5-40°C