Overview

The carbon sulfur analysis instrument CS-206 has the advantages of good reliability, high cost-effectiveness, fast analysis speed, low operating costs, and simple operation and maintenance. This series of instruments adopts pulse heating melting infrared detection method and electronic fluid control technology, widely used in metallurgy, casting, machinery, metal processing, battery materials and other fields. It can quickly and accurately determine the carbon and sulfur content in various materials such as ferrous metals, non-ferrous metals, alloys, carbides, rare earth metals, and other inorganic/organic substances.

Product Details

Carbon and sulfur analysis instrumentScope of application:
Various solid raw materials such as ferrous metals, non-ferrous metals, alloys, carbides, rare earth metals, ores, etc.
Analysis scope:
Carbon: 0.00001% to 10%
Sulfur: 0.00001% to 5%
All of the above can be expanded by changing the sample size to expand the analysis range.

Analysis time: 20-40s

sample weight

Generally, it is 0.1-0.5g, and the sample size can be adjusted according to the sample content

Carrier gas: high-purity oxygen

Power gas: ordinary nitrogen (or purified compressed air) 0.3-0.5Mpa

Power supply: AC220V, 50Hz, 5kVA

Instrument Features:

  • Simultaneously obtaining the content of carbon and sulfur elements in one analysis;

  • Effectively detect carbon and sulfur content at the ppm level;

  • High precision mass flow meter (MFC), with precise flow control and high analysis accuracy; The infrared detection pool adopts the NDIR detection principle, with a wide range and stable operation. The high-frequency induction furnace can quickly and uniformly heat and burn;

  • Furnace head constant temperature heating;

  • Multi channel oxygen supply design for higher stability;

  • Automatic cleaning and ash removal, multi-stage filtration dust removal;

  • Optional high-efficiency physical water removal device;

  • Optional oxygen purification module.

Analysis principle

After purification, pure oxygen enters the combustion chamber and is heated by induction heating in a high-frequency furnace. The carbon (C) and sulfur (S) in the sample are converted into CO2 and SO2 under oxygen rich conditions. The generated CO2 and SO2 enter the corresponding optical detection unit for detection after passing through a dust removal and water purification device. The detection signal can be processed to obtain the percentage content of carbon and sulfur elements. The residual gas containing CO2, SO2, and O2 is discharged through a dedicated pipeline after passing through an absorption device.

碳硫分析仪器

The core component infrared detection pool adopts the NDIR detection principle that has been tested for decades, with high technological maturity, large range span, and stable operation.


  • Selecting efficient and long-life platinum rhodium alloy precious metal micro infrared light sources and metal reflectors for higher optical efficiency;

  • Can effectively detect carbon/sulfur content at the ppm level;

  • Selecting efficient narrowband infrared filters with good rectangular coefficients, high cut-off depth, high characteristic wavelength transmittance, small bandwidth, and high center wavelength positioning accuracy can better shield the interference of other element spectral lines.

Accurate fluid control technology


  • The flow controller adopts a high-precision mass flow meter (MFC) as the main control module, replacing traditional mechanical flow controllers, with more accurate flow control and higher analysis accuracy.

Automatic cleaning and ash discharge design


  • Automatic positive pressure pulse dust removal after analysis, improving the stability of raw material analysis and reducing equipment maintenance work;

  • Multi stage filtering and dust removal design ensures that the infrared detection chamber is not contaminated and eliminates the trouble of cleaning the chamber for users.

Furnace head constant temperature heating device

  • Improve the stability and accuracy of analysis for low content samples and sulfur content.