Product Overview:
Although iron is an essential trace element for the human body. Iron itself is not toxic, but excessive intake or ingestion of iron products can also lead to iron poisoning. Acute iron poisoning mostly occurs in children, while chronic iron poisoning mostly occurs in middle-aged and elderly people over 45 years old, with males being the majority. According to reports, iron poisoning can also induce epilepsy.
This monitoring device is capable of long-term unmanned automatic continuous operation according to on-site settings, and is widely applicable to industrial pollution source discharge wastewater, industrial process wastewater, and other occasions. According to the complexity of the on-site testing conditions, corresponding preprocessing systems can be selected to ensure the reliability of the testing process, accurate testing results, and fully meet the on-site requirements of different occasions.
Product principle:
This product is determined by spectrophotometric method. Under certain acidity conditions, the ferrous ions in the sample form a red complex with the indicator. The monitor detects the color change and converts it into iron value for output. The amount of colored complex generated is equivalent to the amount of iron.
This method is applicable to wastewater containing iron in the range of 0-50mg/L.
Technical Specifications:
serial number |
Specification Name |
Technical specifications and parameters |
1 |
Scope of Application |
This product is suitable for wastewater containing iron in the range of 0-50mg/l |
2 |
test method |
O-phenanthroline colorimetric method |
3 |
measurement range |
0~50mg/l |
4 |
detection limit |
0.05 |
5 |
resolution |
0.001 |
6 |
accuracy |
± 10% or ± 0.1mg/l (whichever is greater) |
7 |
repetitiveness |
10% or 0.1mg/l (whichever is greater) |
8 |
zero drift |
±0.05mg/l |
9 |
range drift |
±10% |
10 |
measurement period |
Small testing cycle of 20 minutes, adjustable |
11 |
sampling period |
Time interval (adjustable), hourly or trigger measurement mode, adjustable |
12 |
calibration cycle |
Automatic calibration (adjustable from 1 to 99 days), manual calibration can be set according to actual water samples |
13 |
maintenance cycle |
The maintenance interval is greater than one month, about 30 minutes each time |
14 |
reagent consumption |
Generally less than 0.8 yuan/sample |
15 |
human-machine operation |
Touch screen display and command input |
16 |
Self checking protection |
Self diagnosis of instrument working status, data will not be lost after abnormalities or power outages After abnormal reset or power outage, the instrument automatically eliminates residual reactants and resumes operation automatically |
17 |
data storage |
Data storage for no less than six months |
18 |
input interface |
discrete signal |
19 |
output interface |
2 RS232 digital outputs, 1 4-20mA analog output |
20 |
working environment |
For indoor work, it is recommended to maintain a temperature of 5-28 ℃ and a humidity of ≤ 90% (without condensation) |
21 |
Power supply and power consumption |
AC230±10%V,50~60Hz,5A |
22 |
size |
Height 1500 x Width 550 x Depth 450 (mm) |