What are the precautions for the electric oxygen permeability tester
Date: 2025-07-02Read: 14
useElectric oxygen permeability testerWhen, it is necessary toSample preparation, equipment operation, environmental control, safety protection, data managementStrictly control key processes to ensure the accuracy, repeatability, and safety of test results. The following are specific precautions:1、 Sample preparation and installation precautions
1. Sample size and integrity
-
Size matching:
- The sample diameter should be ≥ the inner diameter of the test chamber sealing ring (usually ≥ 50mm), ensuring that the effective testing area (such as 50 cm ²) is unobstructed;
- Avoid burrs and curls on the edges of the sample, as this may result in poor sealing or scratches on the testing chamber.
-
Defect inspection:
- Inspect the surface of the sample for defects such as pinholes, cracks, wrinkles, etc. using an optical microscope (≥ 100x) or a transparency meter;
- For high barrier materials such as aluminum foil composite films, special attention should be paid to the uniformity of heat sealing edges or coatings.
-
thickness consistency:
- Use a micrometer to measure the thickness of the sample, ensuring that the thickness deviation of the same batch of samples is ≤ 5%, to avoid calculation errors in transmittance due to uneven thickness.
2. Sample pretreatment
-
cleaning treatment:
- Remove oil stains, fingerprints, dust and other pollutants from the surface of the sample (can be wiped with a dust-free cloth dipped in isopropanol);
- Avoid using chlorine containing cleaning agents to prevent corrosion of metal testing chambers.
-
State adjustment:
- Equilibrate the sample under testing temperature and humidity conditions (such as 23 ℃/50% RH) for at least 4 hours to eliminate the influence of stress on the barrier performance;
- For water containing materials (such as hydrogel coating film), it is necessary to soak them in advance to water absorption saturation state.
3. Installation and sealing
-
Sealing ring selection:
- Select the appropriate sealing ring based on the sample material (such as silicone rubber suitable for -60 ℃~200 ℃, and fluororubber suitable for chemical corrosion resistance scenarios);
- Regularly replace the aging sealing ring (recommended every 500 tests or 3 months) to avoid leakage and low data.
-
Installation strength:
- Use a torque wrench to evenly tighten the test chamber bolts (refer to the equipment manual for torque values, such as 1.5-2.0 N · m);
- Avoid excessive tightening that may cause deformation of the sample or damage to the sealing ring.
-
leak check:
-
After installation, conduct an air tightness test (such as filling with 0.1 MPa nitrogen gas, holding pressure for 5 minutes, and a pressure drop of ≤ 0.001 MPa to be qualified);
-
Leakage can cause oxygen to seep through gaps, leading to inflated test results.
2、 Equipment operation precautions
1. Test parameter settings
-
Standard selection:
- Select corresponding standards based on product type (such as ASTM D3985, ISO 15105-2, GB/T 19789) to avoid parameter confusion and result incomparability;
- For example, food packaging commonly uses ASTM D3985, while drug packaging must comply with USP<671>.
-
Temperature and Humidity Control:
- The testing temperature should be consistent with the actual usage environment of the sample (such as selecting 4 ℃ for cold chain drugs and 38 ℃ for baked goods);
- Humidity has a significant impact on hydrophilic materials such as nylon and EVOH, and requires precise control (such as 50% RH ± 2%);
- Avoid temperature fluctuations greater than 1 ℃/h, otherwise it may cause thermal expansion and contraction of the sample, leading to seal failure.
-
gas flow rate:
- It is recommended to set the flow rate of the upper chamber carrier gas (N ₂) to 20-50 mL/min to ensure that oxygen is fully carried to the sensor;
- The flow rate of the lower chamber mixture (O ₂/N ₂) needs to be stable (such as 100 mL/min) to avoid concentration fluctuations affecting the current signal.
2. Calibration and Verification
-
Standard membrane calibration:
- Use Mocon certified standard film (such as OTR=0.5 cm ³/(m ² · 24h · 0.1MPa)) for calibration, and it is recommended to calibrate every 3 months;
- Before calibration, the standard membrane needs to be equilibrated under test conditions for 2 hours to eliminate the influence of temperature on transmittance.
-
blank test:
- Regularly conduct no sample testing (only installing sealing rings) to verify the background noise of the equipment (current fluctuations should be less than 0.01 pA);
- If the background noise exceeds the standard, it is necessary to check the sensor connection or clean the testing chamber.
3. Testing process monitoring
-
Real time data recording:
- Monitor parameters such as current, temperature, humidity, etc., and draw real-time curves (such as current time curves);
- During normal testing, the current should gradually stabilize. If there is a continuous increase or fluctuation greater than 5%, the machine should be stopped immediately to inspect the sample or equipment.
-
Termination condition judgment:
-
Set termination conditions according to standards (such as stable current for 10 minutes or reaching the preset test time);
-
Avoid premature termination that leads to inaccurate data or excessive testing that wastes resources.
3、 Environmental and safety precautions
1. Laboratory environment requirements
-
Temperature and Humidity Control:
- It is recommended to maintain a laboratory temperature of 20 ℃~25 ℃ and a humidity of ≤ 60% to avoid moisture or condensation on electronic components of the equipment;
- If the environmental humidity is too high, a dehumidifier or air conditioner should be equipped.
-
Ventilation and explosion prevention:
- When testing high concentrations of oxygen (such as>25% O ₂), it is necessary to operate in a fume hood to prevent the risk of explosion caused by oxygen accumulation;
- The equipment must comply with ATEX explosion-proof standards (such as Ex d IIB T4), and the electrical control box should be kept away from the testing chamber.
-
Electromagnetic interference isolation:
- Avoid placing equipment near strong electromagnetic sources such as large motors and frequency converters to prevent interference with current signals;
- Suggest using shielded cables to connect the sensor to the host.
2. Gas management safety
-
Gas cylinder fixation and labeling:
- Oxygen cylinders need to be fixed with dedicated brackets to prevent tipping over;
- Paste warning signs such as "high-purity oxygen" and "flammable and explosive" on the surface of the gas cylinder to avoid accidental operation.
-
Pressure reducing valve and pipeline inspection:
- Regularly check the outlet pressure of the pressure reducing valve (recommended 0.2~0.5 MPa) to avoid damaging the sensor due to excessive pressure;
- Use stainless steel or PTFE tubing to avoid rubber tube aging and leakage;
- Check the pipeline joints for air leaks every month with soapy water.
-
Gas Switching Specification:
-
When replacing gas, the inlet valve of the equipment should be closed first, and the residual gas in the pipeline should be emptied before replacing the gas cylinder;
-
Avoid mixing air (containing 21% O ₂) to prevent impure concentration of the test gas.
4、 Notes on Data Management and Maintenance
1. Data recording and traceability
-
Raw data preservation:
- Save the original test data (including current, temperature, humidity, timestamp) to avoid only recording the final results;
- It is recommended to use the device's built-in software to automatically generate PDF reports and backup them to the cloud or external hard drive.
-
Audit tracking function:
- If the device supports FDA 21 CFR Part 11, the audit trail function must be enabled to record information such as the operator and parameter modification time;
- Regularly check audit logs to ensure that data cannot be tampered with.
2. Equipment maintenance and upkeep
-
Daily cleaning:
- Wipe the inner wall of the testing chamber with a soft cloth every week to avoid dust accumulation that may affect the sensitivity of the sensor;
- Clean the sealing ring with alcohol swab every month to prevent leakage caused by aging and cracking.
-
Consumables replacement cycle:
-
sealing ring: Replace every 500 tests or 3 months;
-
desiccantReplace every 3 months (or based on color indicator discoloration);
-
Electrochemical sensorReplace according to the service life (usually 2-3 years) or output attenuation (such as current drop>20%).
-
Fault handling:
-
If the device alarms' sensor failure ', it is necessary to immediately shut down and contact the manufacturer for repair to avoid damaging the equipment during forced testing;
-
Common fault codes (such as E-01 indicating air path blockage and E-02 indicating temperature exceeding limit) need to be recorded and analyzed for their causes.